Determination of selected endogenous anabolic androgenic steroids and ratios in urine by ultra high performance liquid chromatography tandem mass spectrometry and isotope pattern deconvolution
Impacte
Scholar |
Altres documents de l'autoria: Pitarch-Motellón, Jorge; Sancho, Juan V; Ibáñez, Maria; Pozo, Óscar J.; Roig-Navarro, Antoni F.
Metadades
Mostra el registre complet de l'elementcomunitat-uji-handle:10234/9
comunitat-uji-handle2:10234/33596
comunitat-uji-handle3:10234/33597
comunitat-uji-handle4:
INVESTIGACIONMetadades
Títol
Determination of selected endogenous anabolic androgenic steroids and ratios in urine by ultra high performance liquid chromatography tandem mass spectrometry and isotope pattern deconvolutionAutoria
Data de publicació
2017-09Editor
ElsevierCita bibliogràfica
PITARCH-MOTELLÓN, Jorge, et al. Determination of selected endogenous anabolic androgenic steroids and ratios in urine by ultra high performance liquid chromatography tandem mass spectrometry and isotope pattern deconvolution. Journal of Chromatography A, 2017, vol. 1515, p. 172-178.Tipus de document
info:eu-repo/semantics/articleVersió de l'editorial
http://www.sciencedirect.com/science/article/pii/S0021967317311482Versió
info:eu-repo/semantics/draftParaules clau / Matèries
Resum
An isotope dilution mass spectrometry (IDMS) method for the determination of selected endogenous anabolic androgenic steroids (EAAS) in urine by UHPLC–MS/MS has been developed using the isotope pattern deconvolution ... [+]
An isotope dilution mass spectrometry (IDMS) method for the determination of selected endogenous anabolic androgenic steroids (EAAS) in urine by UHPLC–MS/MS has been developed using the isotope pattern deconvolution (IPD) mathematical tool. The method has been successfully validated for testosterone, epitestosterone, androsterone and etiocholanolone, employing their respective deuterated analogs using two certified reference materials (CRM). Accuracy was evaluated as recovery of the certified values and ranged from 75% to 108%. Precision was assessed in intraday (n = 5) and interday (n = 4) experiments, with RSDs below 5% and 10% respectively. The method was also found suitable for real urine samples, with limits of detection (LOD) and quantification (LOQ) below the normal urinary levels. The developed method meets the requirements established by the World Anti-Doping Agency for the selected steroids for Athlete Biological Passport (ABP) measurements, except in the case of androsterone, which is currently under study. [-]
Proyecto de investigación
Generalitat Valenciana (Research group of excellence Prometeo II 2014/023 and Collaborative Research on Environment and Food Safety ISIC/2012/016) ; University Jaume I (PB1-1B2013-55)Drets d'accés
© 2017 Published by Elsevier B.V.
http://rightsstatements.org/vocab/InC/1.0/
info:eu-repo/semantics/openAccess
http://rightsstatements.org/vocab/InC/1.0/
info:eu-repo/semantics/openAccess
Apareix a les col.leccions
- IUPA_Articles [309]