Development and validation of a method to determine thiabendazole and o-phenylphenol in wastewater using micellar liquid chromatography-fluorescence detection
Impacte
Scholar |
Altres documents de l'autoria: Vives-Peris, Vicente; Roca Genovés, Pascual; Tayeb Cherif, K.; Esteve-Romero, Josep
Metadades
Mostra el registre complet de l'elementcomunitat-uji-handle:10234/9
comunitat-uji-handle2:10234/7013
comunitat-uji-handle3:10234/8638
comunitat-uji-handle4:
INVESTIGACIONAquest recurs és restringit
http://dx.doi.org/10.1002/elps.201500580 |
Metadades
Títol
Development and validation of a method to determine thiabendazole and o-phenylphenol in wastewater using micellar liquid chromatography-fluorescence detectionData de publicació
2016-10Editor
WileyCita bibliogràfica
PERIS VICENTE, Juan; ROCA GENOVÉS, Pasqual; TAYEB CHERIF, K.; ESTEVE ROMERO, Josep. Development and validation of a method to determine thiabendazole and o-phenylphenol in wastewater using micellar liquid chromatography-fluorescence detection. Electrophoresis (2016), v. 37, issue 19, pp. 2517-2521Tipus de document
info:eu-repo/semantics/articleVersió de l'editorial
http://onlinelibrary.wiley.com/doi/10.1002/elps.201500580/epdfVersió
info:eu-repo/semantics/publishedVersionParaules clau / Matèries
Resum
A micellar liquid chromatographic method to determine thiabendazole (TBZ) and o-
phenylphenol in wastewater is described here. The sample was directly injected without
any additional treatment other filtration. The ... [+]
A micellar liquid chromatographic method to determine thiabendazole (TBZ) and o-
phenylphenol in wastewater is described here. The sample was directly injected without
any additional treatment other filtration. The pesticides were resolved in ⬍11 min, us-
ing a mobile phase of 0.10 M SDS—6% 1-pentanol buffered at pH 3 running through a
C18 column at 1 mL/min. The detection was performed by fluorescence at 305/360 and
245/345 nm excitation/emission wavelengths for TBZ and o-phenylphenol, respectively.
The method was validated following the directives of the Validation and Peer Review
of U.S. Environmental Protection Agency Chemical Methods of Analysis guidelines in
terms of selectivity, quantitation range (0.01–0.02 to 2 mg/L), detection limit (0.005–
0.008 mg/L), trueness (92.1–104.2%), precision (⬍13.9%), robustness (⬍6.6%), and sta-
bility under storage conditions. The procedure was applied to the screening of TBZ and
o-phenylphenol in wastewater samples from citrus packing plants, agricultural gutters,
urban sewage, as well as in influent and effluent wastewater treatment plants. [-]