Patient-specific simulation of the intrastromal ring segment implantation in corneas with keratoconus
Impacto
Scholar |
Otros documentos de la autoría: Lago, M.A.; Rupérez, M.J.; Monserrat, C.; Martínez-Martínez, F.; Martínez-Sanchis, S.; Larra, E.; Díez-Ajenjo, M.A.; Peris-Martínez, C.
Metadatos
Mostrar el registro completo del ítemcomunitat-uji-handle:10234/9
comunitat-uji-handle2:10234/7035
comunitat-uji-handle3:10234/8617
comunitat-uji-handle4:
INVESTIGACIONEste recurso está restringido
http://dx.doi.org/10.1016/j.jmbbm.2015.07.023 |
Metadatos
Título
Patient-specific simulation of the intrastromal ring segment implantation in corneas with keratoconusAutoría
Fecha de publicación
2015Editor
ElsevierISSN
1751-6161; 1878-0180Cita bibliográfica
LAGO, M. A., et al. Patient-specific simulation of the intrastromal ring segment implantation in corneas with keratoconus. Journal of the mechanical behavior of biomedical materials, 2015, vol. 51, p. 260-268.Tipo de documento
info:eu-repo/semantics/articleVersión de la editorial
http://www.sciencedirect.com/science/article/pii/S1751616115002647Palabras clave / Materias
Resumen
Purpose
The purpose of this study was the simulation of the implantation of intrastromal corneal-ring segments for patients with keratoconus. The aim of the study was the prediction of the corneal curvature recovery ... [+]
Purpose
The purpose of this study was the simulation of the implantation of intrastromal corneal-ring segments for patients with keratoconus. The aim of the study was the prediction of the corneal curvature recovery after this intervention.
Methods
Seven patients with keratoconus diagnosed and treated by implantation of intrastromal corneal-ring segments were enrolled in the study. The 3D geometry of the cornea of each patient was obtained from its specific topography and a hyperelastic model was assumed to characterize its mechanical behavior. To simulate the intervention, the intrastromal corneal-ring segments were modeled and placed at the same location at which they were placed in the surgery. The finite element method was then used to obtain a simulation of the deformation of the cornea after the ring segment insertion. Finally, the predicted curvature was compared with the real curvature after the intervention.
Results
The simulation of the ring segment insertion was validated comparing the curvature change with the data after the surgery. Results showed a flattening of the cornea which was in consonance with the real improvement of the corneal curvature. The mean difference obtained was of 0.74 mm using properties of healthy corneas.
Conclusions
For the first time, a patient-specific model of the cornea has been used to predict the outcomes of the surgery after the intrastromal corneal-ring segments implantation in real patients. [-]
Publicado en
Journal of the mechanical behavior of biomedical materials, 2015, vol. 51Derechos de acceso
Copyright © Elsevier
http://rightsstatements.org/vocab/InC/1.0/
info:eu-repo/semantics/restrictedAccess
http://rightsstatements.org/vocab/InC/1.0/
info:eu-repo/semantics/restrictedAccess
Aparece en las colecciones
- EMC_Articles [797]