Xanthine derivatives quantification in serum by capillary zone electrophoresis
Impact
Scholar |
Other documents of the author: Vives-Peris, Vicente; Rambla Alegre, María; Durgbanshi, Abhilasha; Bose, Devasish; Esteve-Romero, Josep; Marco Peiró, Sergio
Metadata
Show full item recordcomunitat-uji-handle:10234/9
comunitat-uji-handle2:10234/7013
comunitat-uji-handle3:10234/8638
comunitat-uji-handle4:
INVESTIGACIONThis resource is restricted
http://dx.doi.org/10.1093/chromsci/bmt164 |
Metadata
Title
Xanthine derivatives quantification in serum by capillary zone electrophoresisAuthor (s)
Date
2015-06-22xmlui.dri2xhtml.METS-1.0.item-edition
© The Author [2013]. Published by Oxford University Press. All rights reserved.ISSN
0021-9665Another identifiers
PERIS-VICENTE, Juan, et al. Xanthine Derivatives Quantification in Serum by Capillary Zone Electrophoresis. Journal of chromatographic science, 2014, 52.9: 1121-1126.Type
info:eu-repo/semantics/articlePublisher version
http://chromsci.oxfordjournals.org/content/52/9/1121Version
info:eu-repo/semantics/publishedVersionSubject
Abstract
A capillary electrophoresis method was developed to quantify caffeine and theophylline, xanthine derivatives with bronchodilator activity. Buffer concentration, pH and applied voltage were optimized using a central ... [+]
A capillary electrophoresis method was developed to quantify caffeine and theophylline, xanthine derivatives with bronchodilator activity. Buffer concentration, pH and applied voltage were optimized using a central composite design-face centred. Separation conditions were: silica capillary tube, 75 μm (i.d.) and 61 cm (total length); absorbance detection, 280 nm; borate buffer, 20 mM, pH 9.0; applied voltage, 25 kV and 1 psi injection/8 s. Validation was performed in blank serum following the International Conference Harmonization guidelines: resolution (peaks without overlapping), linear range (0.125–50 µg/mL; r2 > 0.9999), limits of detection and quantification (10; 20 and 33; 66 ppb for caffeine and theophylline, respectively), intra- and inter-day precision (Relative standard deviation lower than 1.9%) and accuracy (98–101%). Migration times were <8 min. This method is simple, specific and suitable and reaches high label claims (98.7–100.4%) in pharmaceutical formulations analysis. Moreover, the method was applied to the monitoring of the analytes in serum of patients. [-]
Is part of
Journal of chromatographic science, 2014, 52.9: 1121-1126Rights
© The Author [2013]. Published by Oxford University Press. All rights reserved.